Multiskan Spectrum

  Multiskan Spectrum



The Multiskan Spectrum spectrophotometer offers a high number of capabilities in one instrument. It provides unlimited wavelength selection, measurement at low UV to visible wavelengths and both cuvette and microplate reading capabilities. SkanIt Software allows easy assay setup and powerful data handling. Moreover, the Multiskan Spectrum has been designed for easy integration with automated systems. With its excellent feature set, the Multiskan Spectrum is ideal for a wide range of photometric applications, such as nucleic acid and protein analysis, enzyme assays, cytotoxicity and cell proliferation assays as well as apoptosis assays.
   Product Detail


Unlimited wavelength selection for easy swapping of assays
Any wavelength from 200 nm to 1000 nm in 1 nm steps can be selected without ever having to change a filter due to the built-in monochromator of the Multiskan Spectrum. Sample absorbances can be scanned over the whole spectrum either in a microplate or in cuvettes. 

Flexible sample formats from cuvette to screening assays
For maximum flexibility, Multiskan Spectrum accepts all microplate formats from 6 to 384 wells, as well as cuvettes from standard to ultramicro capabilities. The cuvette reader has two ports, allowing the user to measure both the sample and reference. Multiple photometric readings per well can be taken on large volume cell culture plates. Even simultaneous use of the microplate and cuvette is possible.  With the pathlength correction function, the results obtained using the microplate can be directly compared with data from any other spectrophotometer.

Logical assay setup using SkanIt Software
Powerful SkanIt Software allows easy assay setup, effortless data reduction, and comprehensive report formatting. The unique step-list feature makes assay setup visual and flexible and the workflow logical and easy to follow. There are two editions available: a Research Edition for scientists working with Life Science Research and a Drug Discovery Edition offering features needed for compliance with FDA’s 21 CFR Part 11 for the Drug Discovery Industry.

IQ/OQ/PQ package for verifiable performance
An instrument qualification IQ/OQ/PQ package is available for the Multiskan Spectrum to verify the optical performance of the instrument.

 

 Multiskan Spectrum
 Product# Price  Description  
       
 51118600      -    Multiskan Spectrum with cuvette, SkanIt Software Drug Discovery Edition   SELECT  
 51118650      -    Multiskan Spectrum with cuvette, SkanIt Software Research Edition   SELECT  
 51118700      -    Multiskan Spectrum without cuvette, SkanIt Software Drug Discovery Edition   SELECT  
 51118750      -    Multiskan Spectrum without cuvette, SkanIt Software Research Edition   SELECT  
 51870301      -    SkanIt Software for Multiskan Spectrum, Research Edition   SELECT  
 51870401      -    SkanIt Software for Multiskan Spectrum, Drug Discovery Edition   SELECT  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Wavelength range :200 -1000 nm
Read-out range :0-4 Abs
Linearity :0-3 Abs, ±2.0% at 450nm
Accuracy :±1% or ±0.005 Abs (0-2 Abs)
±2% (2-3 Abs)
Precision :SD<0.005 Abs or CV < 1% (0-2 Abs)
CV < 2% (2-3 Abs)
Optical bandpass :2 nm
Wavelength accuracy :±1 nm
Stray light :<0.02% at 230 nm
Microplate types :6-, 12-, 24-, 48-, 96- and 384-well plates
Cuvette types :Two cuvettes, for sample and reference From standard to ultramicro: glass, plastic or quartz
Incubator temperature range :Ambient (+25°C) +4°C to 45°C